To set the integrated volume, press the "SELECT
Button" to make the "Integrated volume" setting
value blink, and set the desired integrated volume
with the "INCREASE Button" and "DECREASE
Button". Press and hold to make fast forward.
[Setting range]
0.010
~
900L
Max.: 30 hours × Instantaneous flow setting value
※Setting exceeding 900 L is not
allowed.
※Setting
increments
0.010
~
9.990L:0.01L
10.00
~
99.90L:0.1L
100.0
~
900.0L:1L
※When the
instantaneous flow and the integrated
volume are set, the sampling time is automatically
calculated and the value is displayed at the "Time"
position.
※Set to "
" for continuous sampling.
Setting procedure
When 0.010L is displayed, press "DECREASE
Button" once.
When 900.0 is displayed, press "INCREASE
Button" once.
(When "Integrated volume” is set to "
", "
" is also displayed at the "Time" position)
Press
↓
(Example) Integrated volume: 1 L
↓
Press
or
⑤
To set the standby time, press "SELECT Button" to
make the "Standby" setting value blink, and set the
desired standby time with the "INCREASE Button"
and "DECREASE Button". Press and hold to make
fast forward.
[setting range] 0:00 to 24:00 (0 min to 24 h)
※The setting
increment is one-minute.
Press
↓
(Example) Standby time: 5 min
↓
Press
or