7.4 Setting of the Flow Rate Conversion Temperature of the Calibrator
Set the conversion temperature of the pump at the time of flow rate calibration within the range of 0
to 40°C (in 1°C steps, Factory setting: 25°C). By using this function, the pump can use the same
conversion temperature with the calibrator during calibration.
△
Note
If a calibrator without pressure compensation is used, adjust it to 101.3kPa.
①
(If it is not in the calibration mode), activate the
calibration mode and make sure that "25°C
tStd" is displayed. (When the conversion
temperature is set to 20°C, "20°C tStd" is
displayed.)
②
Press "MODE Button" once to check that "25°C
tCAL" is displayed. (For example, if the
conversion temperature is set to 15 °C, “15 °C
tCAL” is displayed)
Press once
↓
③
Press "START/STOP Button".
Press
④
Check that "tCAL" is blinking.