40
Press the "MODE Button" to exit "Leak Check" without making a judgement during leak
check (
⑤
above). In this case, "
" is displayed.
7.7 Flow Calibration Procedures
Operate the pump under the condition that the flow rate is around 300mL/min with no load to
calibrate the flow rate.
①
Connect the calibrator to the pump. See the figure below.
Example of device connections
②
(If it is not in the calibration mode), activate the
calibration mode and make sure that "25°C tStd" is
displayed. (When the conversion temperature is set
to 20°C, "20°C tStd" is displayed.)
③
Press " MODE Button" four times and check that
"CLF" is displayed.
Press four times
↓