12.
Replace the outer O- ring of the holder.
13.
Insert the window holder including the new O- ring into the
flow cell.
14.
Adjust the holder so that the mounting holes of the plate
match the mounting threads of the flow cell.
Important note:
Please clean the sealing surfaces carefully before
reassembling the unit. If necessary use lubrication grease to
guarantee proper sealing.
Do not damage or sheer the O- rings.
Pic
tu
re
1
0
Pic
tu
re
11
15.
Put the new O- ring into the groove of the window holder.
16.
Put the window into the holder.
Pic
tu
re
1
2
Pic
tu
re
1
3
17.
Put a new O- ring into the optic holder.
Important note:
This O- ring does not have a sealing function; it protects
the window against damages.
18.
Put the optic holder over the window and screw it by using the
mounting screws.
Pic
tu
re
1
4
Important note:
Please make sure that the O- rings stay inside their grooves
during this procedure. We strictly recommend a 30-minute
pressure test under process conditions (without the optical
components) to guarantee proper sealing. This procedure will
avoid possible damages in case of leakage.