Replacement of gaskets
In
tro
du
ct
io
n
Ignoring of the following hints will cause a loss of warranty
Qualified technical personnel must perform repair and maintenance.
Before beginning any work the sensor must be cleaned and flushed thoroughly. Depending on
customers’ application product residues can be very dangerous (aggressive, poisonous). Please
handle the system very carefully due to possible leakage etc.
Avoid pulling and twisting of the lamp cable.
Avoid applying excessive force during assembling and disassembling of the sensor.
Screw in all bolts and cable glands stalwart only.
Please work carefully during the replacement of the measurement gaskets.
The lamp replacement should be done in a dry and clean location, to protect the optical
components from dirt.
Please be sure that no dust or other particles penetrates into the optical assembly.
Do not touch the lenses.
In case of any dirty component please clean carefully by using fresh water and a cloth (lint-
free), dry all components by using instrument air.
Use suitable tools only.
Required tools:
-
2 x open end wrench - 22 mm
-
Flat blade screwdriver - 2 mm
-
Hexagonal spanner – 1.5 mm
-
Hexagonal spanner – 2.5 mm
-
Hexagonal spanner – 3.0 mm
-
Small bowl to drop components
Pic
tu
re
1
1.
Please loose the upper hexagonal nut of the cable gland by
using a 22 mm open-end wrench. Use the second
22 mm open-end wrench to hold the lower hexagonal nut
of the cable gland in position. This will assure that you do
not twist the lamp cable.
2.
Remove Allen screws and / or air purge connectors.
Pic
tu
re
2