Screw off lamp / detector arm.
4.
Shift the lamp / detector arm until you have access to the
electrical/cable connections.
5.
Loosen all wires of the lamp cable by using the 2 mm
flat blade screwdriver.
6.
Remove lamp arm and cable carefully.
Pic
tu
re
4
Pic
tu
re
5
7.
Loosen Allen screws (3x) by using the
1,5 mm hexagonal spanner.
8.
Remove lamp assembly carefully.
9.
Loosen Allen screws (6pc.
M4 x 16 [DIN 912]
) by using
the 3 mm hexagonal spanner.
Pic
tu
re
6
Pic
tu
re
7
10.
Carefully remove sapphire window out of the window
holder.
Important note:
If the window sticks in the holder, remove the holder first
(Step No. 11) after that carefully push the window out of it’s
sealing.
11.
Remove the window holder carefully out of the flow cell.
Important note:
If the holder sticks inside the flow cell, please use the mounting
screws to push the holder out of its sealing. Please make sure that
the holder does not tilt during this procedure.
Do not apply excessive force during this procedure; be careful not
to damage the sealing surfaces for the holder or flow cell.
Pic
tu
re
8