4.8.2.3
Fixed Temperature and Pressure
The calculated concentrations are compensated for both temperature and pressure.
Typically, this temperature and pressure compensation uses the live values from the
analyzer’s cell temperature and pressure measurement sensors. However, there is also
the capability to use static values of temperature and pressure for this compensation. To
use a
Fixed Temperature
and / or a
Fixed Pressure
value for compensation of
concentration values, use the F4 / F3 keys to highlight the checkbox of the fixed value to
be used, then press
↵
to place an X in the box.
4.8.2.4
Sample and Zero Sample Rate
The values shown in the
Sample Rate
and
Zero Sample Rate
fields determine how
frequently the analyzer updates the displayed concentrations and air demand when the
analyzer is in sampling mode (Sample Rate) and in back purge mode (Zero Sample rate).
The time values entered in these two fields are in units of milliseconds. The default value
for each sampling rate is 500 ms.
4.8.3 Display
The
Display
sub-panel, shown in Figure 39, allows the user to configure certain parameters
that affect
how data is displayed on the analyzer’s local display screen.
Figure 39: Display Sub-Panel
4.8.3.1
Analysis 1 and 2 Trend Graph Ranges
The displayed x and y axes for the trend graphs shown on the Analysis 1 and Analysis 2
panels can be configured by the user on this panel using the
Analysis 1/2 X Range
and
Analysis 1/2 Y Range
data entry fields. The Y ranges are concentrations, affecting the
displayed vertical ranges on the Air Demand trend on the
Analysis 1
panel and the
Component concentration trends on the
Analysis 2
panel. These scales can be adjusted
if any of the trend lines are found to be commonly going off of the currently displayed
Summary of Contents for 943-TGX-CE
Page 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Page 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Page 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Page 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Page 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Page 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...