Table 7: Keypad Grey Key Functions
Key Label
Function
F1
Changes the active panel to the previous panel. The previous panel is the
tab to the left of the currently displayed tab. When at the left-most tab,
pressing this key will move to the right-most tab.
F2
Changes the active panel to the next panel. The next panel is the tab to
the right of the current displayed tab. When at the right-most tab, pressing
this key will move to the left-most tab.
F3
On a panel with actionable controls or editable fields, pressing this key
will move to the previous (top to bottom, right to left) control or field.
F4
On a panel with actionable controls or editable fields, pressing this key
will move to the next (top to bottom, left to right) control or field
F5
Returns to the Analysis 1 panel immediately from anywhere within the
Local Display User Interface
F6
Pressing this key while the analyzer is in sampling mode will switch the
analyzer to back-purge mode. Pressing this key while the analyzer is in
back-purge mode, provided that the Status relay state is Normal, will place
the analyzer in sampling mode. This key is functional on any panel of the
local display user interface.
NOTE: F6 is obtained by pressing
the ↑Shift key at the same time as
F1
←
Deletes the last numerical value. Only functional when editing a numerical
data entry field.
Esc
Leaves a data entry field when editing without saving the current value.
Only functional when editing a numerical data entry field.
↵
(ENTER)
After highlighting an editable numerical data entry field using the F4 / F3
keys, pressing
↵
will allow the user to edit the data. After editing the
numerical data using the numerical keys, pressing
↵
again will save the
changes to the analyzer.
4.3
Analysis 1 Panel
The
Analysis 1
panel shown in Figure 17 shows important current analytical parameter results,
status indicators, and operator controls.
4.3.1 Online / Offline Mode Toggle
The
Online / Offline Mode Toggle
at the top right of the Analysis 1 panel shown in Figure
19 is a button that can be used to toggle between
Online
and
Offline
modes.
Figure 19: Online / Offline Mode Toggle
When this toggle is set to Online, the indicator will be green. In this case, the Control relay
will be in the Auto (green) state, indicating that the analyzer believes the data being output
on the analyzer’s analog outputs to be valid. Prior to performing maintenance or other tasks
Summary of Contents for 943-TGX-CE
Page 132: ...Revision 4 132 April 2020 Figure 81 Oven Cabinet Door Removed ...
Page 133: ...Revision 4 133 April 2020 Figure 82 Control Cabinet Door Removed ...
Page 134: ...Revision 4 134 April 2020 Figure 83 Power Steam Air Signals Connection Details ...
Page 135: ...Revision 4 135 April 2020 Figure 84 AC Wiring Schematic ...
Page 136: ...Revision 4 136 April 2020 Figure 85 DC Signals and Wiring Diagram ...
Page 137: ...Revision 4 137 April 2020 Figure 86 Flow Diagram ...