Revision 4
104
April 2020
Figure 69: Replacing the Anti-Solarant Solution
1. Disconnect the source optical fibre from the fluid holder bracket on the UV Lamp
Supply (Item #1 in Figure 69). Let the fibre hang freely and ensure that the end does
not contact anything.
It is suggested that a protective plastic cap is placed onto the end of the
fibre. This fibre is removed from the light path to protect it from unfiltered
exposure to the UV Source Lamp).
2. Loosen the light shield retaining screw on the bottom right-side of the fluid holder and
remove the light shield. (#4 in Figure 69)
3. Loosen the fluid holder locking screw (s) on the left-side of the fluid holder using a
7/64" ball driver.(#2 in Figure 69)
4. Lift the fluid holder from the base by grasping the fluid holder filler plug (#3 in Figure
69).
5. Remove the filler plug from the top of the fluid holder.
6. Dump out the old anti-solarant solution.
The Anti-Solarant Solution is not hazardous to the environment.
1
2
3
4
Summary of Contents for 943-TGX-CE