12
[6] Attach the SpO
2
probe.
There are several types of SpO
2
probe depending on the monitoring
purpose. Select an appropriate type according to the intended use.
The lightweight type such as SD-5C or TG-26(UD-5C) is suitable for
the LX-5630.
SR-5C
TG-25(SP-5C)
TG-26(UD-5C)
for adult, reusable
for adult, reusable
for universal use
SD-5C(25) for adult, semi-disposable
CE-5C for earlap
SD-5C(20) for pediatric, semi-disposable
NOTE
: The accuracy of this instrument, like that of all other dual-wavelength
oximeters, can be influenced by the presence of abnormal
hemogrobins such as carboxyhemoglobin (HbCO) and methemoglbin.
The tables below show the errors which may occur due to these
hemoglobins. The instrument may be also affected by cardiogreen or
intravascular dyes.
HbCO
METHEMOGLOBIN
SpO
2
1%
5%
10%
1%
5%
10%
50%
-0.1%
-0.7%
-1.5%
0.2%
1.3%
3.2%
70%
-0.1%
-0.7%
-1.5%
-0.6%
-2.3%
-3.2%
90%
-0.2%
-0.8%
-1.6%
-1.5%
-6.0%
-9.6%
100%
-0.2%
-0.8%
-1.7%
-1.8%
-7.5%
-12.2%