[2] Turn the power switch to "ON".
When the power is turned ON, the SpO
2
measurement mode is in
continuous mode. To change the mode to intermittent mode, press the
mode switch. In the intermittent mode, the SpO
2
will be measured at
30 seconds interval and n o pulse waveform is sent, but ECG and
respiration will be continuously measured. In the continuous SpO
2
measurement mode, the SpO
2
value and pulse waveform will be
continuously measured.
CAUTION
In the intermittent mode, the SpO
2
is measured at 30 seconds interval.
Even when the SpO
2
value suddenly changes during this interval, the
transmitter will only show the last SpO
2
measurement value. For the
patient with the possibility of sudden change, select the continuous
mode.
In the intermittent mode, the battery life is about 5 days and about 3
days in continuous mode.