18
n
CE-5C (Earlap Probe) Attachment
The CE-5C is suitable for long-term measurement and is used by
attaching to an ear.
Before attaching the probe to a patient, clean the skin using a cloth
moistened with sterilizing alcohol.
CE-5C
(1) Attach the ear holder to the probe cable.
(2) Place the ear hanger to the ear.
•
It is possible to slide the ear
hanger up and down.
•
The ear hanger can be used on
both small and large ear.
•
It can be also turned.
(3) Attach the ear clip to the ear.
Before clipping, massage or warm the
earlap for good circulation.
Inset the tube here.
Insert the tube (at 16.5cm from the
probe end) to the ear holder.
Push in the hook part to
the notch.
A clip to fix on to a
bed sheet or a breast
Ear Clip
Ear Holder
Ear Hanger
Ear
Hanger