26
No.
Date
Checker
Installation Place
Unit LX-5630
S/No.
Purchase Date
Items
Details of th e Check
Criteria
Judgement
Appearance
Visually check for any damage,
cracks, chink, chips and peeled
nameplate on the housing.
No abnormality
should be found.
c
OK
c
NG
Battery
Compartment
Visually check the connect ing
spring inside the device and
battery cap.
Spring should not be
transformed,
deformed or rusted.
c
OK
c
NG
Power
Turn ON/OFF power to verify
proper switch operation.
With battery
installed, the LCD
will display.
c
OK
c
NG
ECG
Connector
Visually check for connector of
main unit and patient cable.
No scratch ,chips,
dust should be
found.
c
OK
c
NG
Patient Cable
Visually check the wire coating of
patient cable.
No cracks, kinks, or
damage should be
found.
c
OK
c
NG
SpO
2
Relay
Cable
Visually check the cable sheath
and connectors.
No cracks, kinks, or
damage should be
found.
c
OK
c
NG
SpO
2
Sensor
Visually check for cable sheath
sensor and connectors.
No cracks, kinks,
dust, damage should
be found.
c
OK
c
NG
Wireless
Channel
Visually verify the transmit
channel ID, and group code.
Follow instructions of the
wireless channel manager for the
receiver.
Must correspond
with wireless
channel check list.
c
OK
c
NG
Operational
function
Turn ON the power and make
sure the operation is normal.
Waveform is
received without any
problem.
c
OK
c
NG
Display
function
Turn ON the power and verify the
SpO
2
data, bargraph, etc. is
properly displayed.
All data should be
properly displayed.
c
OK
c
NG
Periodic
Check
Check the date of previously
performed periodic check.
Should be within
one month.
c
OK
c
NG