5. Receiver Channel Setup
Select the receiver channel at the patient monitor corresponding to the
telemetry transmitter. The channel number will be shown on the
display. Also, if the group code is set on the transmitter, it should be
also set on the receiver. (For details of channel setup and group code
setup, refer to the operation manual for the patient monitor on the
receiver side.)
If the set channel number does not correspond to the actually received
channel number, the monitor will display a cautionary waveform as
shown below. This function will prevent interference with other
transmitters or external sources.
NOTE
: This function will work only if the receiver incorporates the
interference detection function.
Example of the DS-5800N telemetry patient's display when interference
is present.
This cautionary waveform will also be displayed when a mismatch on
group code or channel number occurs.
CAUTION
This radio frequency device is susceptible to interference from outside
sources. Depending on receiving signal condition, the waveform may
include noise, false pacer spike, etc. To prevent these interferences,
it is recommended to install and use the telemetry antenna system.