14
n
TG-25 (SP-5C) (Finger Clip Probe) Attachment
The TG-25 (SP-5C) is suitable for short-term measurement, and should
be clipped to a fingertip. It should be positioned so that the LED (light
source) is at the base of the fingernail.
Light Source
Sensor
Proper attachment
Finger is not sufficiently
inserted.
Finger is inserted too far.
Measurement cannot be taken.
CAUTION
Precautions for Use of TG-25 (SP-5C)
·
Cover the probe with a black cover if it is exposed to direct sunlight. Some
measurement error may occur if the probe is exposed to direct sunlight during
measurements.
·
The TG-25 (SP-5C) is designed for use on adult finger. Do not use on other
parts.
·
When removing the probe, do not pull by holding the cable.
·
The TG-25 (SP-5C) is designed for short-term measurement. If measurements
are taken continuously over a long period of time, change the measuring finger
every 2 hours to prevent low-temperature burn. Especially for continuous use on
patient with peripheral circulatory disturbance, change the measuring finger
more frequently.
·
Do not use a tape to fix the probe. It may cause edema or congestion.
·
The probe should be positioned so that the light source is at the base of the
fingernail. Take care not to insert a finger too deep to prevent injury.
·
It may not be possible to take measurements in case of an excessively high
hematocrit value. In such case, reattach the probe to a thinner finger.
·
Remove nail polish before taking measurements to enable stable measurements.
·
Before attaching the probe to a patient, clean the probe using a cloth moistened
with sterilizing alcohol.
·
The connector terminals of the probe should be protected against water or
chemicals.
·
It may not be possible to take measurements in place with vibrations or while
walking. In such case, use the TG-26 (UD-5C) or SD-5C for more stable
measurements.
·
Inaccurate measurements can be caused by the placement of SpO
2
probe on an
extremity with a blood pressure cuff, arterial catheter, or intravascular line
.
TG-25 (SP-5C)