114
Questions and Troubleshooting
5.
Other Functions
(1) Method to hide messages on [READY] screen
• Change the fiber image from X/Y view to X magnified view or Y magnified view by
pressing X/Y key.
(2) Too many repetitions until “Test Finish” indicated in [Arc Calibration]
• The splicer needs to repeat the arc calibration after replacing the electrodes or when the
environmental conditions change drastically (page 87). The number of arc calibrations
can be set to a specific amount. When the splicer completes the set amount of
calibrations, it indicates “Test Finish”. However, this does not mean it is completely
calibrated.
(3) “Test Finish” is never indicated after many repetitions in [Arc Calibration]
• Execute [Stabilize Electrodes] function in [Maintenance Menu]. If the splicer still does
not indicate “Test Finish”, replace the electrodes. See [Replace Electrodes] section.
(4) No arc power change after [Arc Calibration]
• An internal factor is calibrated and adjusted for the specific arc power selected. The
displayed arc power in each splice mode does not change.
• The calibration results affect all splice modes.
(4) Method to input different comments after each splice in splice results data
• See page 79 [Splice Memory Comment].
(5) Method to download splice results from splicer to PC
• Contact your distributor.