100
Maintenance Menu
• Cleave angle threshold does not link to the parameter "Cleave Limit" in
splicing modes. Cleave angle threshold is independently set for arc calibration.
See page 84 [Splice Settings] to change cleave angle threshold.
“Good” message
Arc power and splicing position calibration are successfully completed. Press Escape
key to exit.
“Not Adequate” message
Arc power and splicing position calibration are completed but further calibration is
strongly recommended, as the change from the previous arc calibration is too large.
Press Enter key to perform arc calibration, or Escape key to exit even though arc
calibration is not completed.
• In some cases, multiple iterations of arc calibration are needed until the calibration
process is successfully completed and the "Test Finish" message is displayed. Arc
calibration can be considered almost completed if multiple iterations are completed
without receiving the message.
• Number threshold can be set so that "Test Finish" message is displayed after
specific number of arc calibrations are performed. See page 84 [Splice Settings] for
detail.
Select the [Arc Calibration]
in the [Maintenance Menu
Arc Calibration Result
Arc Calibration
Enter key
Enter key