INF-TN2FSC-E
83
10.2.1 “Logger Operation” mode
There are two logging modes, i.e., quick logger that permits operation from the measurement
screen and logger that is set from the menu screen. Logger is of two different modes, i.e.,
“CONTINUOUS” mode and “SET TIME” mode.
• Quick logger
• Logger
(1) “CONTINUOUS” mode
(2) “SET TIME” mode
Logger data will be stored at the selected site name in the Site Memory.
(1) Quick logger
The quick logger starts when quick logger start button
in the measurement screen is pressed.
When memory card is not inserted, cursor does not move to function button.
The quick logger exits when a period of an hour has elapsed since it was started or when quick
logger stop button is pressed.
• Logging time: 1 hour, fixed
• Cycle: 10 seconds, fixed
• Measured data type: 3 types (unit and number of digits after decimal point are the same) dis-
played in measurement screen and status display
In the case of 3-line or 2-line display, only the flow rate on the first line
is stored.
Start button
Measurement screen
(2) Logger
(1) “CONTINUOUS” mode
“CONTINUOUS” mode is the mode to perform logging in a fixed period from start date and
time to exit date and time.
Exit occurs upon elapse of exit date and time or when the stop button is pressed. As the exit
time varies by the start time and period, there are cases where it is different from the set exit
time.
• Logging time: Start date and time to exit date and time
• Cycle: 10 seconds to 24 hours
• Measured data type: 14 measured data types and status display