10.1.5 TOTALIZER: when performing the total process of measured data
(totalize)
(1) Total start/stop from Menu screen
[Operation]
(1) Select “TOTALIZER” on the SITE SETUP screen by
pressing the or key.
Press the
ENT
key to display the total output selection
screen.
Total process and setting of total output can be performed.
To set the Total start/stop, there are two modes that Quick logger operated from the Measurement
screen and setting from Menu Screen, operation from Measurement screen is given priority.
(3) Move the cursor to “START” by the or key,
and then press the
ENT
key.
The total will start.
(2) Pressing the
ENT
key enables you to select the total
output mode.
Select the mode from MANUAL, FIXED TIME, and
TIMER by pressing the or key.
(For details of MODE, refer to description of the
mode in the following page.)
(4) To stop the total output in the middle, move the cur-
sor to “STOP” and press the
ENT
key.
“STOP” is also possible from the totalizing calcula-
tion display button on the measurement screen.
Total will not start unless the cursor is pointed to “START”
and the
ENT
key is pressed.
Even if the timer setting is performed after scheduled Start
date may pass than the time of the main unit, total will not
be carried out.
CAUTION