To test the weighing scale one has to put a (known) weight into one pan of the collecting vessel and click “
Test
Scale
” (blue gear wheel icon) in the tab “
Scale
”.
To simulate precipitation amounts to check the output signals of the pulse and analogue outputs, one has to
select the desired amount from the drop-down menu in the tab “
Simulation
” and click “
Simulate Precipitation
”
(cloud icon). The analogue and pulse outputs will return signals according to the applied settings in the
rain[e]
Setup.
9
SDI-12
Interface
The communication using SDI-12 protocol via SDI-12 interface is based on the ‘SDI-12 A Serial-Digital
Interface Standard for Microprocessor-Based Sensors, Version 1.3, 2012’. The
RAINE-HYDRO
can be used in
bus mode parallel to other
RAINE-HYDRO
s.
The following subset of SDI-12 commands were implemented into the
RAINE-HYDRO
.
For further details to the SDI-12 protocol we recommend the afore-mentioned standard document or the website
www.SDI-12.org.
Implemented SDI-12 commands:
Command
Function
Answer of the sensor
a!
Acknowledge Active
a<CR><LF>
?!
Address Query Command
a<CR><LF>
a
I
!
Send Identi
fi
cation
allccccccccmmmmmmvvvxx...xx<CR><LF>
aAb!
Change Address
b<CR><LF>
aM!
Start Measurement
atttn<CR><LF>
aMC!
Start Measurement and Request CRC
atttn<CR><LF>
aC!
Start Concurrent Measurement
atttnn<CR><LF>
aCC!
Start Concurrent Measurement and Request
CRC checksum
atttnn<CR><LF>
aD0!
aD1!
Send Data (Pu
ff
er 0)
Send Data (Pu
ff
er 1)
a<values<CR><LF>
a<values><CRC><CR><LF>
resp. with CRC checksum
aM1!
Generate Variance
atttn<CR><LF>
aM2!
Generate Heating Data
atttn<CR><LF>
aM3!
Start Intensity Measurement Average Value,
Maximum and Minimum
atttn<CR><LF>
aMC1!
Generate Variance and Request CRC
atttn<CR><LF>
aMC2!
Generate Heating Data and Request CRC
atttn<CR><LF>
aMC3!
Start Intensity Measurement Average Value,
Maximum and Minimum and Request CRC
atttn<CR><LF>
aC1!
Generate Variance
atttnn<CR><LF>
aC2!
Generate Heating Data
atttnn<CR><LF>
aC3!
Start Intensity Measurement Average Value,
Maximum and Minimum
atttnn<CR><LF>
aCC1!
Generate Variance and Request CRC
atttnn<CR><LF>
aCC2!
Generate Heating Data and Request CRC
atttnn<CR><LF>
aCC3!
Start Intensity Measurement Average Value,
Maximum and Minimum and Request CRC
atttnn<CR><LF>
aV!
Start Veri
fi
cation
atttn<CR><LF>
a = address of the respective sensor; standard sensor address = 0
17
SDI-RAINE-HYDRO
USER MANUAL
700-SDI-RAINE-HYDRO Rev1 03 Mar 2021
Part# 21184