The Expert Con
fi
guration screen is an easy to use tool to completely customise the con
fi
guration of the
rain[e]
output signals. Options are visible depending on the selected output modes. The following list shows the available
options, sub-options and value ranges.
Pulse Output (non-isolated) / (isolated)
• Operating Mode
◦
Pulse
►
Closing Time
10...500 ms in steps of 5 ms
►
Resolution
0.01...1 mm in steps of 0.01 mm
◦
Rain YES / NO
Precipitation Event (non-isolated) / (isolated)
►
Amount
0.10...1 mm in steps of 0.10 mm
►
Start Time
20...60 s in steps of 1 s
►
End Time
20...600 s in steps of 1 s
Serial Port
• Operating Mode
◦
SDI-12
►
Average Time
1...60 min in steps of 1 min
SDI-12 - Address
0...9, A...Z, a...z
Load Cell Adjustment
Fig. 15
The Load Cell Adjustment screen may be used to adjust the load cell. It will give the number of cycles the program
has carried out, the mean value of the measured weight and the variance of the values. If a calibration weight is
used, a calibration factor will be given.
To adjust the load cell
fi
rst select the COM port of the
RAINE-HYDRO
from the drop-down menu. If desired, put
a calibration weight in the collecting vessel and enter the weight into the corresponding text box. To start the
adjustment click the green Start Adjustment button.
Pressing the button with the red ‘X’ will stop the running adjustment.
Adjustment has to done again if the deviations in the diagnostics are higher than ± 30 mg.
15
SDI-RAINE-HYDRO
USER MANUAL
700-SDI-RAINE-HYDRO Rev1 03 Mar 2021
Part# 21184