Generation of the data: precipitation intensity,
average value, max. / min. intensity of the last x
minutes*.
Measured variables
Unit
Bu
ff
er 0
Average value over x minutes*
mm/min
Max. intensity of the last x minutes* mm/min
Min. intensity of the last x minutes*
mm/min
The default value for the period x is 10 min. The
value can be changed with the
rain[e]
Com-
mander.
aM3!
a0033<CR><LF>
aC3!
a00303<CR><LF>
with CRC
checksum
aMC3!
a0033<CR><LF>
aCC3!
a00303<CR><LF>
* The acquisition of these values starts with the command and they have to be requested after exactly x minutes
with the aD0! command.
Start Veri
fi
cation - aV! (Error string)
For service purposes the command
aV!
can be used to do a system analysis and generate an error string. The
command
aV!
has the same format as the command
aM!
(see above). The
RAINE-HYDRO
answers to the
command
aV!
with ‘a0039’.
Syntax
Command
Answer
aV!
a0039<CR><LF>
a
– Sensor address
a
– Sensor address
V
– Command ‘Start Veri
fi
cation’
003
– Seconds the sensor needs until the
measured
data
can
be
returned
(=
3
s)
9
– Number of provided data
!
– End of command
<CR><LF>
– End of answer
Example:
Command
Answer
1V!
10039<CR><LF>
The measured data can be requested with the command
aD0!
. (see
Send Data
).
Data output
Value range
Bu
ff
er 0
For internal use
0...99
Error due to exceeding of the heating temperature by 10° C
0 or 1
Error heating
0 or 1
Error interior temperature sensor
0 or 1
Error funnel temperature sensor
0 or 1
For internal use
0 or 1
For internal use
0 or 1
For internal use
0 or 1
For internal use
0 or 1
+0 = ok; +1 = error
The data will not be overwritten until the next
C
,
M
, or
V
command and can be read several times until then.