Calibration Adjustment
4.6 Final Calibration
4
4-13
4.6.7 Gain Ohm
Proceed as follows to do the Gain Ohm calibration:
1.
Press
to select calibration adjustment step
Gain Ohm (CL 0860):IDLE
2.
Connect the UUT to the 5500A as shown in Figure 4-8.
Notice that the sense leads must be connected directly to the test tool.
ST8003.WMF
Figure 4-8. Four-wire Ohms calibration connections
3.
Set the 5500A to the first test point in Table 4-4. Use the 5500A “COMP 2 wire”
mode for the calibration adjustments up to and including 100 k
Ω
. For the higher
values, the 5500A will turn off the “COMP 2 wire” mode.
4.
Set the 5500A to operate (OPR).
5.
Press
to start the calibration.
6.
Wait until the display shows the calibration status
:READY
.
7.
Press
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points.
8.
When you are finished, set the 5500A to Standby.
9.
Continue at Section 4.6.8.
Table 4-4. Ohm Gain Calibration Points
Cal Step
Input Value
Gain Ohm (CL 0860) [Cap. Pos. (CL 0920), Cap.Neg. (CL 0921)]
1)
100
Ω
Gain Ohm (CL 0861) [Cap. Pos. (CL 0922), Cap.Neg. (CL 0923)]
1)
1
k
Ω
Gain Ohm (CL 0862) [Cap. Pos. (CL 0924), Cap.Neg. (CL 0925)]
1)
10
k
Ω
Gain Ohm (CL 0863) [Cap. Pos. (CL 0926), Cap.Neg. (CL 0927)]
1)
100
k
Ω
Gain Ohm (CL 0864)
1 M
Ω
Gain Ohm (CL 0865) [Gain Ohm (CL 0866)]
2)
10
M
Ω
1)
The capacitance measurement current calibrations (Cap.Pos. and Cap.Neg) are done automatically after
the Gain Ohm calibration
.
2)
The Gain Ohm (CL0866) calibration step is done automatically after the Gain Ohm (CL0865) calibration.
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......
Page 30: ...Fluke 123 124 125 Service Manual 2 16...
Page 32: ......
Page 62: ......
Page 64: ......
Page 80: ......
Page 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Page 88: ...Fluke 123 124 125 Service Manual 5 10...
Page 90: ......