Performance Verification
3.5 Input A and Input B Tests
3
3-19
Table 3-3. Volts AC Measurement Verification Points
Sensitivity
(Oscilloscope)
Range
1)
(Meter)
5500A output
Volts rms
5500A
Frequency
Reading A-B
200 mV/div
500 mV
500 mV
60 Hz
494.0 to 506.0
500 mV
20 kHz
486.0 to 514.0
2V/div
5V
5V
20 kHz
4.860 to 5.140
5V
60 Hz
4.940 to 5.060
20V/div
50V
50V
60 Hz
49.40 to 50.60
50V
20 kHz
48.60 to 51.40
1)
The 500V and 1250V range will be tested in Section 3.5.14
3.5.11 Input A and B AC Input Coupling Test
Proceed as follows to test the Input A and B AC coupled input lower transition point:
1.
Connect the test tool to the 5500A as for the previous test (see Figure 3-5).
2.
Select the following test tool setup:
•
Use the setup of the previous step (AUTO time base, traces at vertical center).
•
Using
select 200 mV/div for Input A and B (500 mV range).
•
Fluke 123/124: press
/
to open the SCOPE INPUTS menu, and choose:
INPUT
A:
AC |
NORMAL | INPUT B:
AC | NORMAL
•
Fluke 123/124: press
/
to open the SCOPE INPUTS menu.
•
Fluke 123/124: press
to open the TRIGGER menu, and choose:
INPUT: A | SCREEN UPDATE:
FREE RUN | AUTO RANGE:
> 1Hz
•
Fluke 125: press
to open the A MEASUREMENTS menu, and choose:
•
Fluke 125: press
to open the A INPUT menu, and choose:
INPUT
A:
AC |
NORMAL
•
Fluke 125: press
to open the B MEASUREMENTS menu, and choose:
•
Fluke 125: press
to open the B INPUT menu, and choose:
INPUT
B:
AC | NORMAL
•
Fluke 125: press
to open MENU.
•
Fluke 125: press
to open the TRIGGER menu, and choose:
UPDATE:
FREE RUN | AUTO RANGE:
> 1Hz
3.
Set the 5500A to source an AC voltage, to the first test point in Table 3-4 (NORMAL
output, WAVE sine). Set the 5500A to Operate (OPR).
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......
Page 30: ...Fluke 123 124 125 Service Manual 2 16...
Page 32: ......
Page 62: ......
Page 64: ......
Page 80: ......
Page 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Page 88: ...Fluke 123 124 125 Service Manual 5 10...
Page 90: ......