Fluke 123/124/125
Service Manual
4-6
4.4 Contrast Calibration Adjustment
After entering the Maintenance mode, the test tool display shows
Warming Up (CL 0200):IDLE (valid)
.
Do not press
now! If you did, turn the test tool off and on, and enter the
Maintenance mode again.
Proceed as follows to adjust the maximum display darkness (CL0100), the default
contrast (CL0110) , and the maximum display brightness (CL0120).
1.
Press
a three times to select the first calibration step. The display shows:
Contrast (CL 0100) :MANUAL
2.
Press
CAL. The display will show a dark test pattern, see Figure 4-2
3.
Using
adjust the display to the maximum darkness, at which the test pattern
is only just visible.
4.
Press
to select the default contrast calibration. The display shows:
Contrast (CL 0110) :MANUAL
5.
Press
CAL. The display shows the test pattern at default contrast.
6.
Using
set the display to optimal (becomes default) contrast.
7.
Press
to select maximum brightness calibration. The display shows:
Contrast (CL 0120) :MANUAL
8.
Press
CAL. The display shows a bright test pattern.
9.
Using
adjust the display to the maximum brightness, at which the test
pattern is only just visible.
10.
You can now :
•
Exit, if only the Contrast had to be adjusted. Continue at Section 4.7.
OR
•
Do the complete calibration. Press
to select the next step (Warming Up),
and continue at Section 4.5.
Figure 4-2. Display Test Pattern
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......
Page 30: ...Fluke 123 124 125 Service Manual 2 16...
Page 32: ......
Page 62: ......
Page 64: ......
Page 80: ......
Page 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Page 88: ...Fluke 123 124 125 Service Manual 5 10...
Page 90: ......