Fluke 123/124/125
Service Manual
3-24
7.
When you are finished, set the 5500A to Standby
Table 3-7. V DC and V AC High Voltage Verification Tests
Sensitivity
(Scope)
Range
(Meter)
5500A
output Vrms
5500A
Frequency
Main (DC)
Reading A-B
Secondary (AC)
Reading A-B
200V/div 500V
0V
DC
-000.5
to
+000.5
+500V DC +497.0
to
+503.0
-500V
DC
-497.0 to -503.0
500V
60Hz
494.0 to 506.0
500V
10 kHz
486.0 to 514.0
500V/div
1250V
600V
10 kHz
0.570 to 0.630
600V
60Hz
0.584 to 0.616
+600V DC +0.592
to
+0.608
-600V
DC
-0.592 to -0.608
0V DC -0.005
to
+0.005
3.5.16 Resistance Measurements Test
Proceed as follows:
1.
Connect the test tool to the 5500A as shown in Figure 3-9.
ST8003.WMF
Figure 3-9. Test Tool Input A to 5500A Normal Output 4-Wire
2.
Select the following test tool setup:
•
Press
to select auto ranging (
AUTO
in top of display).
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......