Performance Verification
3.5 Input A and Input B Tests
3
3-7
1.
Short circuit the Input A and the Input B shielded banana sockets of the test tool. Use the
BB120 banana to BNC adapter, and a 50
Ω
(or lower) BNC termination.
2.
Select the following test tool setup:
•
Turn Input B on (if not already on).
•
Press
to select auto ranging (
AUTO
in top of display).
(
toggles between
AUTO
and
MANUAL
ranging).
•
Fluke 123/124: press
/
to open the SCOPE INPUTS menu.
•
Fluke 123/124: press
to open the SCOPE OPTIONS menu, and choose :
SCOPE
MODE:
NORMAL | WAVEFORM MODE:
SMOOTH
•
Fluke 125: press
to open MENU.
•
Fluke 125: press
to open the TRIGGER menu, and choose :
UPDATE:
FREE RUN
•
Fluke 125: press
to open the SMOOTH menu, and choose :
WAVEFORM:
SMOOTH
•
Press
to leave the menus.
3.
Using
toggle the time base between 10 ms/div and 5 ms/div.
(the time base ranging is set to manual now, the input sensitivity is still automatic; no
indication
AUTO
or
MANUAL
is displayed).
After changing the time base wait some seconds until the trace has settled.
Observe the Input A trace, and check to see if it returns to the same position after
changing the time base. The allowed difference is
±
0.04 division (= 1 pixel).
Observe the Input B trace for the same conditions.
4.
Using
toggle the time base between 1
μ
s/div and 500 ns/div. After changing the
time base wait some seconds until the trace has settled.
Observe the Input A trace, and check to see if it is set to the same position after changing
the time base. The allowed difference is
±
0.04 division (= 1 pixel).
Observe the Input B trace for the same conditions.
5.
Using
set the time base to 10 ms/div.
6.
Using
toggle the sensitivity of Input A between 5 and 10 mV/div. After
changing the sensitivity wait some seconds until the trace has settled.
Observe the Input A trace, and check to see if it is set to the same position after changing
the sensitivity. The allowed difference is
±
0.04 division (= 1 pixel).
7.
Using
toggle the sensitivity of Input B between 5 and 10 mV/div. After
changing the sensitivity wait some seconds until the trace has settled.
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......
Page 30: ...Fluke 123 124 125 Service Manual 2 16...
Page 32: ......
Page 62: ......
Page 64: ......
Page 80: ......
Page 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Page 88: ...Fluke 123 124 125 Service Manual 5 10...
Page 90: ......