Fluke 123/124/125
Service Manual
2-6
Common Mode Rejection (CMRR)
>100 dB @ DC
>60 dB @ 50, 60, or 400 Hz
Full Scale Reading
5000 counts
Move influence
±
6 counts max.
True RMS Voltages (VAC and VAC+DC)
Ranges
500 mV, 5V, 50V, 500V, 1250V
Accuracy for 5 to 100% of range
DC coupled:
DC to 60 Hz (VAC+DC)
±
(1% +10 counts)
1 Hz to 60 Hz (VAC)
±
(1% +10 counts)
AC or DC coupled:
60 Hz to 20 kHz
±
(2.5% +15 counts)
20 kHz to 1 MHz
±
(5% +20 counts)
1 MHz to 5 MHz
±
(10% +25 counts)
5 MHz to 12.5 MHz
±
(30% +25 counts)
5 MHz to 20 MHz
±
(30% +25 counts), excluding test leads or
probes
AC coupled with 1:1 (shielded) test leads:
60 Hz (6 Hz with 10:1 probe)
-1.5%
50 Hz (5 Hz with 10:1 probe)
-2%
33 Hz (3.3 Hz with 10:1 probe)
-5%
10 Hz (1 Hz with 10:1 probe)
-30%
DC Rejection (only VAC)
>50 dB
Common Mode Rejection (CMRR)
>100 dB @ DC
>60 dB @ 50, 60, or 400 Hz
Full Scale Reading
5000 counts
The reading is independent of any signal
crest factor.
Move influence
±
6 counts max.
Peak
Modes
Max peak, Min peak, or pk-to-pk
Ranges
500 mV, 5V, 50V, 500V, 1250V
Accuracy:
Max peak or Min peak
5% of full scale
Peak-to-Peak
10% of full scale
Full Scale Reading
500 counts
Frequency (Hz)
Ranges
1Hz, 10Hz, 100Hz, 1 kHz, 10 kHz,
100 kHz, 1 MHz, 10 MHz, and 50 MHz
(Fluke 123) or 70 MHz (Fluke 124, 125).
Frequency Range for Continuous Autoset
15Hz (1Hz) to 50 MHz
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......
Page 30: ...Fluke 123 124 125 Service Manual 2 16...
Page 32: ......
Page 62: ......
Page 64: ......
Page 80: ......
Page 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Page 88: ...Fluke 123 124 125 Service Manual 5 10...
Page 90: ......