Characteristics
2.3 Dual Input Meter
2
2-5
Source
A, B, EXT
EXTernal via optically isolated trigger
probe ITP120
(optional accessory)
Sensitivity A and B (Fluke 123)
@ DC to 5 MHz
0.5 divisions or 5 mV
@ 25 MHz
1.5 divisions
@ 40 MHz
4 divisions
Sensitivity A and B (Fluke 124, 125)
@ DC to 5 MHz
0.5 divisions or 5 mV
@ 40 MHz
1.5 divisions
@ 60 MHz
4 divisions
Voltage level error
±
0.5 div. max.
Slope
Positive,
Negative
Video on A
Interlaced video signals only
Modes
Lines, Line Select
Standards
PAL , NTSC, PAL+, SECAM
Polarity Positive,
Negative
Sensitivity 0.6
divisions
sync.
2.2.4 Advanced Scope Functions
Display Modes
Normal
Captures up to 40 ns glitches and displays analog-like persistence
waveform.
Smooth
Suppresses noise from a waveform.
Envelope
Records and displays the minimum and maximum of waveforms
over time.
Auto Set (Connect-and-View
TM
)
Continuous fully automatic adjustment of amplitude, time base, trigger levels, trigger
gap, and hold-off. Manual override by user adjustment of amplitude, time base, or trigger
level.
2.3 Dual Input Meter
The accuracy of all measurements is within
±
(% of r number of counts) from
18
°
C to 28
°
C.
Add 0.1x (specific accuracy) for each
°
C below 18
°
C or above 28
°
C. For voltage
measurements with 10:1 probe, add probe uncer1%.
More than one waveform period must be visible on the screen.
2.3.1 Input A and Input B
DC Voltage (VDC)
Ranges
500 mV, 5V, 50V, 500V, 1250V
Accuracy
±
(0.5% +5 counts)
Turnover
±
12 counts
Normal Mode Rejection (SMR)
>60 dB @ 50 or 60 Hz
±
1%
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......
Page 30: ...Fluke 123 124 125 Service Manual 2 16...
Page 32: ......
Page 62: ......
Page 64: ......
Page 80: ......
Page 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Page 88: ...Fluke 123 124 125 Service Manual 5 10...
Page 90: ......