Fluke 123/124/125
Service Manual
4-8
3.
Set the 5500A to source a 1 kHz fast rising edge square wave (Output SCOPE,
MODE edge) to the first calibration point in Table 4-1.
4.
Set the 5500A in operate (OPR).
5.
Press
to start the calibration.
6.
Wait until the display shows calibration status
READY
.
7.
Press
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points in Table 4-1.
8.
Set the 5500A to source a 1 kHz square wave (Output SCOPE, MODE wavegen,
WAVE square), to the first calibration point in Table 4-2.
9.
Press
to select the first step in Table 4-2.
10.
Press
to start the calibration.
11.
Wait until the display shows calibration status
READY
.
12.
Press
to select the next calibration step, set the 5500A to the next calibration
point, and start the calibration. Continue through all calibration points Table 4-2.
13.
When you are finished, set the 5500A to Standby.
14.
Continue at Section 4.6.2.
Table 4-1. HF Gain Calibration Points Fast
Cal step
5500A Setting
1)
(1 kHz, no 50
Ω
!)
Test Tool Input Signal
Requirements
1)
(1 kHz, t
rise
<100 ns,
flatness after rising edge:
<0.5% after 200 ns)
HFG & FI AB (CL 0600)
10 mV
20 mV
HFG & FI AB (CL 0601)
25 mV
50 mV
HFG & FI AB (CL 0602)
50 mV
100 mV
HFG & FI AB (CL 0603)
100 mV
200 mV
HFG & FI AB (CL 0604)
250 mV
500 mV
HFG & FI AB (CL 0605)
500 mV
1V
HFG & FI AB (CL 0606)
1V
2V
HFG & FI AB (CL 0607)
[HFG & FI A (CL 0608), HFG & FI B (CL 0628)]
2)
2.5V 5V
1)
As the 5500A output is not terminated with 50
Ω
, its output voltage is two times its set voltage
2)
After starting the first step in this table cell, these steps are done automatically.
Summary of Contents for ScopeMeter 123
Page 6: ......
Page 12: ......
Page 16: ......
Page 30: ...Fluke 123 124 125 Service Manual 2 16...
Page 32: ......
Page 62: ......
Page 64: ......
Page 80: ......
Page 82: ...Fluke 123 124 125 Service Manual 5 4 ST8014 EPS Figure 5 1 Fluke 123 124 125 Main Assembly...
Page 88: ...Fluke 123 124 125 Service Manual 5 10...
Page 90: ......