5820A
Operators Manual
7-4
/
63Hz
47Hz
300VA MAX
FUSE
CAUTION
FOR FIRE PROTECTION
REPLACE ONLY WITH A 250V FUSE
OF INDICATED RATING
MAINS SUPPLY
/
120V
100V
/
240V
220V
T2.5A 250V
(SB)
T1.25A 250V
(SB)
CHANGING LINE FUSE
aag067f.eps
Figure 7-1. Accessing the Fuse
7-3.
Cleaning the Air Filter
XW
Warning
To avoid risk of injury, never operate or power the 5820A
calibrator without the fan filter in place.
W
Caution
To avoid possible damage caused by overheating, keep the
area around the fan unrestricted. If the air intake becomes
restricted, the intake air will be too warm, or the filter will
become clogged.
The air filter must be removed and cleaned every 30 days or more frequently if the
calibrator is operated in a dusty environment. The air filter is accessible from the rear
panel of the calibrator.
Summary of Contents for 5820A
Page 10: ...5820A Operators Manual vi ...
Page 14: ...5820A Operators Manual x ...
Page 16: ...5820A Operators Manual 1 2 ...
Page 32: ...5820A Operators Manual 2 2 ...
Page 40: ......
Page 144: ...5820A Operators Manual 6 2 ...
Page 190: ......
Page 212: ...5820A Operators Manual 7 24 ...
Page 214: ...5820A Operators Manual 8 2 ...
Page 230: ...5820A Operators Manual 9 2 ...
Page 234: ...5820A Operators Manual ...
Page 244: ...5820A Operators Manual A 10 ...
Page 245: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 246: ...5820A Operators Manual B 2 ...
Page 248: ...5820A Operators Manual B 4 ...
Page 250: ...5820A Operators Manual C 2 ...
Page 258: ...5820A Operators Manual E 4 ...
Page 264: ...5820A Operators Manual F 6 ...
Page 272: ...5820A Operators Manual 8 ...