Multifunction Calibrator
Wideband Output Flatness Test
85
7. Record the 1 kHz absolute errors at each output level from Table 22 into the
appropriate column in Table 24. Record the flatness errors from Table 23 for
each output level and frequency into the appropriate column in Table 24. Add
the errors and verify that they are within spec for the appropriate time
internal.
Table 21. Wideband Frequency Accuracy Test Record
Frequency ( Hz )
Frequency Measured
Tolerance Limits
10 Hz
99.99 ms to 100.01 ms
100 Hz
9.999 ms to 10.001 ms
300 Hz
299.97 Hz to 300.03 Hz
500 Hz
499.95 Hz to 500.05 Hz
800 Hz
799.92 Hz to 800.08 Hz
900 Hz
899.91 Hz to 900.09 Hz
1 kHz
999.0 Hz to 1.0001 kHz
1.19 kHz
1.189881 kHz to 1.190119 kHz
2.2 MHz
2.19978 MHz to 2.20022 MHz
3.5 MHz
3.49965 MHz to 3.50035 MHz
3.8 MHz
3.79962 MHz to 3.80038 MHz
10 MHz
9.990 MHz to 10.001 MHz
20 MHz
19.998 MHz to 20.002 MHz
30 MHz
29.997 MHz to 30.003 MHz
Table 22. Wideband Accuracy at 1 kHz Test Record
Output
Level
Measured
Error
ppm
Spec
±
ppm
(90-day)
5790A
1-Year Spec
±
ppm
TUR
GB Factor
Test Limit
2.1 V
2238
24
93
1.000
2238
1.0 V
2900
24
121
1.000
2900
0.3 V
2833
38
74.5
1.000
2833
0.1 V
3400
53
64.1
1.000
3400
30 mV
3333
115
29.0
1.000
3333
10 mV
3900
240
16.2
1.000
3900
3 mV
4833
643
7.5
1.000
4833
1 mV
5400
1720
3.14
0.980
5292