AC Voltage Verification Test
The equipment required for this test is listed in Table 14.
Table 14. Equipment Required for AC V Output Level Tests
1. Place the 5790A on top of the Product and connect the equipment as shown
in Figure 15. Other placements of the equipment are fine as long as they can
be connected with the 12 inch cable.
Note
The point of measurement is at the end of the cable and adapter that
connects to the 5790A. It is essential that the cable not be longer
than 12 inches, it can be shorter.
2. Ensure the guard to ground strap on the Product is connected during the
procedure. Also ensure the Product is set to internal Guard (External Guard
OFF). Refresh the EX GRD button on the 5790A by pressing it off and on 3
times, and then leave it on. The green LED is lit when in the ON position. Set
the range on the 5790A to AUTO and connect the Product guard to the
5790A guard using a short heavy lead.
3. Refer to Table 15 and output .600000 mV @1 kHZ and record the reading
under Measured Value.
4. Test the remaining amplitudes and frequencies in Table 15.
5. Verify that the measured values are within the limits given in the Table 15.
1000V RMS MAX
1000V RMS MAX
3V RMS MAX
INPUT 1
INPUT 2
SHUNT
HI
LO
GUARD
GROUND
10V PEAK
MAX
WIDEBAND
7V RMS MAX
SHELL FLOATING
10V PEAK MAX
5730A Calibrator
5790A
AC MEASUREMENT
STANDARD
5790A Measurement Standard
External
Sense
OFF
External
Guard
OFF
External
Guard
ON
hhp310.eps
Figure 15. AC Voltage Test Setup
Equipment Model
AC Measurement Standard
Fluke 5790A
BNC(F) to Dual-Banana Plug Adapter (2 required)
Pomona 1269
Coax Cable - RG-58A/U or RG-58C/U with BNC(M)
Connectors, 12 inch Long