
5730A
Calibration Manual
36
15. To reject the changes, touch
Cancel
. Otherwise, touch
Continue
to accept
and save the changes, and to continue with calibration.
Once the internal steps are complete, a screen similar to this is shown:
Setup Menu
Calibration
Artifact Calibration
Calibration Complete
The largest change was 30% of spec.
22VDC
Would you like to save changes?
Discard
Save
hhp215.eps
16. Touch
Save
to store the results and adjust the Product. If the security
passcode has not already been entered, the Product prompts to do so before
the results are saved. Touch
Discard
to remove the results from the
calibration.
Summary of Contents for 5730A
Page 4: ......
Page 8: ...5730A Calibration Manual iv ...
Page 10: ...5730A Calibration Manual vi ...
Page 104: ...5730A Calibration Manual 94 ...