Multifunction Calibrator
Calibration Procedures
35
8. Reverse the HI and LO connections at the 732B terminals, and push
Continue
on the Product to continue the calibration process.
9. After this part of the calibration procedure is complete, the Product shows the
reference shifts and prompts to continue. Touch
Continue
. The Product
prompts to refer to the manual for the correct connections and enter the first
calibration value.
10. Connect the Product to the 10 k
Ω
standard as shown in Figure 2 and enter
the value of the standard using processes similar to those described for dc
voltage. If the standard is not between 9 k
Ω
and 11 k
Ω
, an error message is
shown. Start over from this point with a different 10 k
Ω
standard. Push
on the Product again to continue.
11. When the internal 10 k
Ω
reference has been characterized, accept or reject
the changes that are about to be made to the calibration constant. To reject
the changes, touch
Cancel
. Otherwise, touch
Continue
to accept and save
the changes. This lets the calibration process continue.
Resistance Standard
Calibrator
hhp028.eps
Figure 2. 742A-1 and 742A-10 k External Calibration Connections
12. Disconnect the 10 k
Ω
standard, and connect the Product to the 1
Ω
standard.
Enter the value of the 1
Ω
standard using processes similarly described for
the other standards. If the standard is not between 0.9
Ω
and 1.1
Ω
, an error
message appears, and the user can start over from this point with another
standard.
13. Push
Continue
to proceed with calibration.
14. When the internal 1
Ω
reference has been characterized, accept or reject the
changes that are about to be made to the calibration constant.