5730A
Calibration Manual
14
5725A Amplifier:
Range
Frequency
(Hz)
Stability
±
1
°
C
[1]
24 Hours
Temperature Coefficient
Adder
Load Regulation
[2]
Distortion
Bandwidth
10 Hz -10 MHz
±
(% output)
10 - 40
°
C
0 - 10
°
C and
40 - 50
°
C
±
(ppm mV)
±
(ppm output) /
°
C
±
(ppm mV) 150 pF
1000 pF
1100 V
40 - 1 k
1 k - 20 k
20 k - 50 k
50 k - 100 k
10 + .5
15 + 2
40 + 2
130 + 2
5
5
10
30
5
5
10
30
10 + 1
90 + 6
275 + 11
500 + 30
0.10
0.10
0.30
0.40
0.10
0.15
0.30
0.40
Notes:
1.
Stability specifications are included in Absolute specification values for the primary specifications.
2.
The 5725A will drive up to 1000 pF of load capacitance. Absolute specifications include loads to 300 pF and 150 pF as shown
under "Load Limits." For capacitances up to the maximum of 1000 pF, add "Load Regulation."
Voltage Range
Maximum Current Limits
Load Limits
2.2 V
[2]
22 V
220 V
50 mA, 0
°
C-40
°
C
20 mA, 40
°
C-50
°
C
>50
Ω
,
1000 pF
1100 V
6 mA
600 pF
5725A Amplifier:
40 Hz-5 kHz
50 mA
1000 pF
[1]
1100 V
5 kHz-30 kHz
70 mA
300 pF
30 kHz-100 kHz
70 mA
[3]
150
pF
Notes:
1.
The 5725A will drive up to 1000 pF of load capacitance. Absolute specifications include loads to 300 pF and 150 pF as shown
under "Load Limits." For capacitances up to the maximum of 1000 pF, add "Load Regulation."
2.
2.2 V Range, 100 kHz-1.2 MHz only: Absolute specifications cover loads to 10 mA or 1000 pF. For higher loads, load regulation
is added.
3.
Applies from 0
°
C to 40
°
C.
Output Display Formats
....................................... Voltage or dBm, dBm reference 600
Ω
.
Minimum Output
................................................... 10 % on each range
External Sense
...................................................... Applicable for 2.2 V, 22 V, 220 V, and 1100 V ranges; 5730A
<100 kHz, 5725A <30 kHz. Specifications are the same as internal
sense.
Settling Time to Published Specifications
Frequency (Hz)
Settling Time (seconds)
10-120 7
>120 5
Notes:
Plus 1 second for amplitude or frequency range change
Plus 2 seconds for 5730A 1100 V range
Plus 4 seconds for 5725A 1100 V range
Overshoot
.............................................................. <10 %
Common Mode Rejection
....................................... 140 dB, dc to 400 Hz