5730A
Calibration Manual
20
DC Current Secondary Performance Specifications and Operating Characteristics
Range
Stability
±
1
°
C
[1]
24 Hours
Temperature
Coefficient
[2]
Compliance
Limits
Burden
Voltage
Adder
[3]
(
±
nA/V)
Maximum
Load for
Published
Specification
[4]
(
Ω
)
Noise
10 - 40
°
C
0 - 10
°
C
and
40 - 50
°
C
Bandwidth
0.1-10 Hz
Bandwidth
10 Hz-10 kHz
pk-pk RMS
±
(ppm
nA)
±
(ppm nA) /
°
C
ppm
output
+ nA
nA
220
μ
A
2.2 mA
22 mA
220 mA
2.2 A
5 + 1
5 + 5
5 + 50
8 + 300
9 + 7
μ
A
1 + 0.40
1 + 2
1 + 20
1 + 200
1 + 2.5
μ
A
3 + 1
3 + 10
3 + 100
3 + 1
μ
A
3 + 10
μ
A
10
10
10
10
3
[5]
0.2
0.2
10
100
2
μ
A
20k
2k
200
20
2
6 + .9
6 + 5
6 + 50
9 + 300
12 + 1.5
μ
A
10
10
50
500
20
μ
A
5725A
±
(ppm
μ
A)
±
(ppm
μ
A)/
°
C
ppm
μ
A
μ
A
11 A
25 + 100
20 + 75
30 + 120
4
0
4
15 + 70
175
Notes:
Maximum output from the calibrator’s terminals is 2.2 A. Specifications for 220
μ
A and 2.2 mA ranges are increased by a factor of 1.3
when supplied through 5725A terminals.
1.
Stability specifications are included in the Absolute specification values for the primary specifications.
2.
Temperature coefficient is an adder to Absolute specifications. It does not apply unless operating more than
±
5
°
C from
calibration temperature.
3.
Burden voltage adder is an adder to Absolute specifications that does not apply unless burden voltage is greater than 0.5 V.
4.
For higher loads, multiply Absolute specification by:
ion
specificat
published
for
load
maximum
load
actual
x
0.1
1
+
5.
The calibrator’s compliance limit is 2 V for outputs from 1 A to 2.2 A. 5725A Amplifier may be used in range-lock mode down to
0 A.
Minimum Output
.................................................. 0 for all ranges, including 5725A.
Settling Time
........................................................ 1 second for
μ
A and mA ranges; 3 seconds for 2.2 A range; 6 seconds
for 11 range; + 1 second for range or polarity change
Overshoot
............................................................. <5 %
Summary of Contents for 5730A