Features
Softkey Menu Trees
3
3-9
Table 3-1. Front-Panel Features (cont.)
The
TC
(Thermocouple) minijack is used for thermocouple simulation during thermometer
calibration, and thermocouple measurements. You must use the correct thermocouple wire and plug
when using this connector. For example, if simulating a type K thermocouple, use type K
thermocouple wire and type K plug for making connections
The
20A
terminal is the source of current output when the 20 A range is selected (3 A - 20 A)
The
AUX (Auxiliary Output)
terminals are used for ac and dc current outputs, the second voltage
output in dual voltage modes, and ohms sense for 2-wire and 4-wire compensated resistance and
capacitance measurements, and RTD simulation
The
GUARD
terminal is always connected internally to the internal guard shield. This shield is tied to
the NORMAL LO signal ground inside the Calibrator unless the
B
key is pressed so that its
indicator is lit
The
NORMAL
(Normal Output) terminals are used for ac and dc voltage, ohms and capacitance
sourcing, and Resistance Temperature Detector (RTD) simulation.
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...