5522A
Operators Manual
4-44
nn108f.eps
3.
Press the softkey FUNDMTL to select the Calibrator front panel terminals for the
fundamental output, either NORMAL or AUX. The harmonic appears on the AUX
terminals.
4.
Press the softkey HARMNIC to enter the desired harmonic (1 to 50), with a
maximum frequency output of 10 kHz. For example, entering the 7th harmonic
(below). When the control display shows the desired value, press
E
.
nn109f.eps
5.
Press
P
one or more times to return to previous menus.
How to Adjust the Phase
When in the dual ac voltage and ac power output modes, you can set the calibrator to
source two signals with adjustable phase difference. All phase adjustments shift the AUX
waveform in relation to the NORMAL waveform. Phase shift adjustments are entered
either as degrees (0 to
±
180.00) or as a power factor (PF). A leading or positive phase
shift causes the AUX waveform to lead the NORMAL waveform; a lagging or negative
phase shift causes the AUX waveform to lag the NORMAL waveform.
The softkey PHASE is available after pressing the WAVE MENUS softkey that appears
when outputting dual ac voltages or ac power (shown below for ac power output)
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...