Remote Operations
How to Set up the RS-232 UUT Port for Remote Control
5
5-19
UUT
5522A Cali
b
rator
Controller
SERIAL 2
TO UUT
Port
RS-232
Port
Modem Ca
b
le
IEEE-4
88
Ca
b
le
gjh046.eps
Figure 5-6. Testing the RS-232 UUT Port via IEEE-488 Port
Complete the following procedure to test RS-232 UUT port operation via the IEEE-488
port using the Win32 Interactive Control utility.
1.
Complete the “IEEE-488 Port Setup Procedure” earlier in this chapter to set up the
Calibrator for GPIB operation.
2.
Complete “Testing the IEEE-488 Port” to prepare the Calibrator IEEE-488 port for
testing. Before the final step, return to this procedure and continue to Step 3 below.
3.
Go to Start then to the Programs menu.
4.
Select "NI-488.2M software for... (your operating system)".
5.
From the NI488.2M software menu, select "Win32 interactive control".
6.
A DOS window opens with a prompt as shown here:
7.
At the prompt, type the following line to activate the IEEE interface card:
<ibdev 0 4 0 10 1 0>
8.
The second number in this line is the primary address of the calibrator. If the address
has been changed from the factory default, change this line accordingly.
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...