SC1100 Oscilloscope Calibration Option
Remote Commands and Queries
10
10-29
Table 10-1. SCOPE Command Parameters (cont.)
Parameter Description/Example
EDGE
Oscilloscope EDGE mode. Programs 25 mV peak-to-peak, 1 MHz, output at the
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns EDGE.
Example: SCOPE EDGE; OUT 0.5 V, 5 kHz
(Edge, 0.5 V peak-to-peak, 5 kHz.)
LEVSINE
Oscilloscope LEVSINE mode. Programs 30 mV peak-to-peak, 50 kHz, output at the
SCOPE BNC, standby if from OFF or previously in standby. FUNC? returns LEVSINE.
Example: SCOPE LEVSINE; OUT 1 V, 50 kHz
(Leveled sine wave, 1 V peak-to-peak, 50 kHz.)
MARKER
Oscilloscope MARKER mode. Programs the period to 1 ms, output at the SCOPE
BNC, standby if from OFF or previously in standby. FUNC? returns MARKER.
Example: SCOPE MARKER; OUT 2 MS
(Marker, period of 2 ms.)
WAVEGEN
Oscilloscope WAVEGEN mode. Programs 20 mV peak-to-peak, square wave, 1 kHz,
no offset, output impedance 1 M
Ω
, standby if from OFF or previously in standby.
FUNC? returns WAVEGEN.
Example: SCOPE WAVEGEN; OUT 1 V, 1 kHz
(Wave Generator, 1 V peak-to-peak, 1 kHz.)
VIDEO
Oscilloscope VIDEO mode. Programs 100 % output (1 V p-p), line marker 10, format
NTSC. FUNC? returns VIDEO.
Examples:
SCOPE VIDEO; OUT 90
(Video, 90 % output)
SCOPE VIDEO; OUT -70
(Video, -70 % output, inverse video)
PULSE
Oscilloscope PULSE mode. Programs 100 ns pulse width, 1.000
μ
s period, 2.5 V
range. FUNC? returns PULSE.
Example:
SCOPE PULSE; OUT 50 ns, 500 ns; RANGE TP8DB
(Pulse, 50 ns pulse width, 500 ns period, 1.5 V range)
MEASZ
Oscilloscope Impedance/Capacitance measurement (MEAS Z) mode. Programs 50
Ω
range. FUNC? returns MEASZ.
Example:
SCOPE MEASZ; RANGE TZCAP
(MEAS Z mode, capacitance range)
Oscilloscope Overload mode. Programs 5 V dc range. FUNC? returns OVERLD.
OVERLD Example:
SCOPE
OVERLD; OUT 7 V; RANGE TOLAC
(Overload, 7 V output, ac range)
SCOPE
Programs the 5520A-SC1100 oscilloscope calibration option hardware, if installed. The
instrument settings are determined by this command’s parameter. Once in SCOPE mode,
use the OUT command to program new output.
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...