5522A
Operators Manual
6-32
*PUD?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Protected User Data query) Returns the contents of the *PUD (Protected User Data)
memory in definite length format.
Response:
#2
nn<nn characters>
Example:
*PUD?
returns
#216CAL LAB NUMBER 1
Returns #2 then 16 then 16 characters of text (including spaces) stored in the nonvolatile
memory.
RANGE?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Range query) Returns the present output ranges. Both the primary output and secondary
outputs are returned. If there is no secondary output,
0
is returned. Dual outputs are noted
with
P
for primary output (front panel NORMAL terminals) and
S
for secondary output
(front panel AUX terminals).
Response:
<primary output>,<secondary output>
Examples:
DC330MV,0
(dc volts 330 mV range)
DC33MA_A,0
(dc current 33 mA range)
AC3_3V,0
(ac volts 3.3 V range)
AC330MA_A,0
(ac current 330 mA range)
R110OHM,0
(ohms
110
Ω
range)
C1_1UF,0
(capacitance
1.1
μ
F range)
TCSRC,0
(temperature
thermocouple
source)
RTD_110,0
(temperature
RTD
110
Ω
range)
DC3_3V_P,DC3A_AS
(dc power 3.3 V, 3 A ranges)
AC330V_P,AC20A_2S
(ac power 330 V, 20 A ranges)
DC330MV_P,DC3_3V_S
(dual dc volts 330 mV, 3.3 V ranges)
AC330V_P,AC3_3V_S
(dual ac volts 330 V, 3.3 V ranges)
Returns the symbolic name of the single or first output, and return the symbolic name of
the second output (0 if there is no second output).
RANGELCK
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Range Lock command) Locks in the present range, or selects auto ranging for dc
voltage and dc current single outputs. The range automatically unlocks if the output
function changes, for example from dc volts to dc current. When RANGELCK is on, this
is equivalent to the softkey range lock showing locked. When RANGELCK is off, this is
equivalent to the softkey range lock showing auto.
Parameter:
ON
(Locks the dc volts or dc current range)
OFF
(Unlocks the dc volts or dc current range for autoranging)
Example:
RANGELCK OFF
Set the range lock off to allow autoranging for dc volts or dc current.
RANGELCK?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Range Lock query) Returns whether or not the preset dc volts or dc current single
output range is locked.
Response:
ON
(range is locked and autoranging is not allowed)
OFF
(range is not locked and autoranging is allowed)
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...