5522A
Operators Manual
4-42
Peak
Period
RMS (70
%
Peak)
nn026f.eps
Figure 4-11. Sine Wave
Triangle Waves
When the wave selection is tri, the triangle wave is present on the calibrator outputs
(Figure 4-12). The variables for the triangle wave are amplitude, frequency, and dc offset
voltage. Whenever a triangle wave is selected, the Output Display indicates amplitudes in
p-p units.
Peak to Peak
nn027f.eps
Figure 4-12. Triangle Wave
Square Wave
When the wave selection is square, a square wave current or voltage signal is present on
the calibrator outputs (Figure 4-13). The variables for the square wave are duty cycle,
amplitude, frequency, and dc offset voltage. Whenever a square wave is selected, the
Output Display indicates amplitude in p-p units. If the calibrator is set for a single voltage
or current output, the duty cycle of the signal can be set through the keypad. To enter a
new duty cycle, press the DUTY CYCLE softkey and up to five numeric keys followed
by
E
. The negative-going edge of the square wave will move based on the duty
cycle setting.
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...