![Fluke 522A/6 Operator'S Manual Download Page 351](http://html1.mh-extra.com/html/fluke/522a-6/522a-6_operators-manual_2310337351.webp)
SC1100 Oscilloscope Calibration Option
Verification Tables
10
10-37
Table 10-3. SC1100 Option DC Voltage Verification at 50
Ω
(cont.)
Calibrator
Mainframe
Output
Measured Value
(V DC)
Reading x correction
Tolerance (V DC)
-109.9 mV
0.0003148 V
499 mV
0.0012875 V
-499 mV
0.0012875 V
2.19 V
0.005515 V
-2.19 V
0.005515 V
6.599 V
0.0165375 V
-6.599 V
0.0165375 V
AC Voltage Verification
Table 10-4. SC1100 Option AC Voltage Verification
(1 M
Ω
output impedance unless noted)
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-year Spec.
(V p-p)
0.001
1000
0.000041
-0.001
1000
0.000041
0.01
1000
0.00005
-0.01
1000
0.00005
0.025
1000
0.000065
-0.025
1000
0.000065
0.11
1000
0.00015
-0.11
1000
0.00015
0.5
1000
0.00054
-0.5
1000
0.00054
2.2
1000
0.00224
-2.2
1000
0.00224
11
1000
0.01104
-11
1000
0.01104
130
1000
0.13004
-130
1000
0.13004
200
mV
100
0.00024
200
mV
1000
0.00024
200
mV
5000
0.00054
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...