![Fluke 522A/6 Operator'S Manual Download Page 361](http://html1.mh-extra.com/html/fluke/522a-6/522a-6_operators-manual_2310337361.webp)
SC1100 Oscilloscope Calibration Option
Verification Tables
10
10-47
Table 10-17. SC1100 Option Leveled Sinewave Verification: Flatness (cont.)
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation
(V p-p)
1-Year Spec.
(V p-p)
0.0099
360
MHz
0.000496
0.0099
390
MHz
0.000496
0.0099
400
MHz
0.000496
0.0099
480
MHz
0.000496
0.0099
570
MHz
0.000496
0.0099
580
MHz
0.000496
0.0099
590
MHz
0.000496
0.0099
600
MHz
0.000496
0.01 10
MHz
na na
0.01
30
MHz
0.00025
0.01
70
MHz
0.00025
0.01
120
MHz
0.0003
0.01
290
MHz
0.0003
0.01
360
MHz
0.0005
0.01
390
MHz
0.0005
0.01
400
MHz
0.0005
0.01
480
MHz
0.0005
0.01
570
MHz
0.0005
0.01
580
MHz
0.0005
0.01
590
MHz
0.0005
0.01
600
MHz
0.0005
0.01
1000
MHz
0.0005
0.025 10
MHz
na
na
0.025
30
MHz
0.000475
0.025
70
MHz
0.000475
0.025
120
MHz
0.0006
0.025
290
MHz
0.0006
0.025
360
MHz
0.0011
0.025
390
MHz
0.0011
0.025
400
MHz
0.0011
0.025
480
MHz
0.0011
0.025
570
MHz
0.0011
Summary of Contents for 522A/6
Page 4: ......
Page 14: ...5522A Operators Manual x...
Page 18: ...5522A Operators Manual xiv...
Page 22: ...5522A Operators Manual 1 2...
Page 46: ...5522A Operators Manual 1 26...
Page 48: ...5522A Operators Manual 2 2...
Page 54: ...5522A Operators Manual 2 8...
Page 56: ...5522A Operators Manual 3 2...
Page 190: ...5522A Operators Manual 6 2...
Page 240: ...5522A Operators Manual 7 2...
Page 260: ...5522A Operators Manual 8 2...
Page 314: ...5522A Operators Manual 9 52...
Page 370: ...5522A Operators Manual 10 56...
Page 413: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes...
Page 414: ...5522A Operators Manual B 2...
Page 416: ...5522A Operators Manual B 4...