Fluke 434-II/435-II/437-II
Users Manual
27-6
Fluke 434-II @ 50Hz
nominal
Fluke 434-II @ 60Hz
nominal
42.5 ... 57.5 Hz
51 ... 69 Hz
0.001 Hz
0.001 Hz
± 0.01Hz
± 0.01Hz
Power Measurement
Range
Resolution
Accuracy
Watt (VA, var)
i430flex-TF
1 mV/A
max 6000 MW
max 2000 MW
0.1 W … 1 MW
0.1 W … 1 MW
± 1% ± 10 counts
± 1% ± 10 counts
Power Factor (Cos
ϕ
/
DPF)
0…1
0.001
± 0.1% @ nominal load
conditions
Energy Measurement
Range
Resolution
Accuracy
kWh (kVAh, kvarh)
i430flex-TF 10x
Depends on clamp scaling and V nominal
± 1% ± 10 counts
Energy loss
i430flex-TF 10x
Depends on clamp scaling and V nominal
± 1% ± 10 counts
Excluding line resistance
accuracy
Harmonics Measurement
Range
Resolution Accuracy
Harmonic order (n)
DC, 1..50 Grouping: Harmonic groups according to IEC 61000-4-7
Inter-Harmonic order
Off, 1..50 Grouping: Harmonic and Interharmonic subgroups according to IEC
61000-4-7
Volts %f
0.0 … 100.0%
0.1%
± 0.1% ± n x 0.1%
Volts %r
0.0 … 100.0%
0.1%
± 0.1% ± n x 0.4%
Volts Absolute
0.0 … 1000 V
0.1 V
± 5% (*)
Volts THD
0.0 … 100.0%
0.1%
± 2.5%
Amps %f
0.0 … 100.0%
0.1%
± 0.1% ± n x 0.1%
Amps %r
0.0 … 100.0%
0.1%
± 0.1% ± n x 0.4%
Amps Absolute
0.0 … 600 A
0.1 A
± 5% ± 5 counts
Amps THD
0.0 … 100.0%
0.1%
± 2.5%
Watts %f or %r
0.0 … 100.0%
0.1%
± n x 2%
Watts Absolute
Depends on clamp scaling and V nominal
± 5% ± n x 2% ± 10
counts
Watts THD
0.0 … 100.0%
0.1%
± 5%
Phase angle
-360º ... +0º
1º
± n × 1º (8)
*) ± 5 % if
≥
1 % of nominal voltage; ± 0.05 % of nominal voltage if < 1% of nominal voltage.