Figure 4-2 Dose Rate Mode:
In "Dose Rate" mode, the rate, by default, is expressed in micro-rem per hour. The option to change
this to Sievert is available as well.
The "Meter" in the display offers a bar chart, visual representation of the Dose Rate. Additionally,
there is provided on the same display a digital (numeric) representation as well.
The expression of units of measure may be set by the user, as detailed above.
The Meter
The meter is a horizontal, metered dose rate bar along a logarithmic axis ranging from 0 μrem/h - 25
mrem/h across the upper portion of the screen, just below the "Status Bar".
If the dose rate rises above the threshold, this is displayed graphically, and consistent with con-
figuration (threshold) annunciation will activate (vibration, LED, beeper or any combination of the
three in keeping with user preference settings).
Digital
Large digits and a dose rate unit in the center of the display
The units are expressed as (μSv/h, Sv/h - adjusts in keeping with order of magnitude of the dose
rate) and reflect user preferences (rem, or Sievert).