23.4.2 Zero waste filling - weigh checked fill
1.
Place an empty vial on the balance.
How the fills are started depends on the selected filling method:
•
If the filling method is set to
time delay between fills
, filling will commence
automatically after an empty vial is placed on the balance.
•
If the filling method is set to
keypad to start each fill
, the
icon is
displayed once an empty vial is placed on the balance. Press
to start
filling.
i
The pump determines there to be an empty vial on the
balance whenever a weight is displayed on the balance
that is within the vial weight tolerance. Refer to "Vial
weight" on page 61 and "Vial weight tolerance" on
page 61.
i
The balance must achieve a stable weight for a calibration
value to be entered. If the balance is taking too long to
stabilise or not stabilising at all, refer to the balance
manufacturers instructions.