When the pump is dispensing or ready to dispense, the screen background colour is
blue.
Dispense icon
Page indicator icon
Batch information (Fill rate = Fills/min)
Figure 26 - Dispensing or ready to dispense screen
Use the up and down keys (
) to change between display pages whilst a batch is
running.
If the
key or
key is pressed at any point, the pump will stop
immediately and a fill interrupt screen will be displayed.
23.3.1 Calibration whilst dispensing
1. If using a balance, tare the balance using the vial to be filled.
2. Press
to access
Calibrate
mode.
3. Enter the calibration value
i.
No connected balance—
Use the numeric keypad. Press
to
confirm.
ii.
Connected balance—
Place the filled vial on the balance. The value
automatically displays. Press
to confirm.