First user batch approval
i
If FDA 21CFR Part 11 compliance is enabled, a password is
required.
When a batch has finished dispensing, the user is prompted to enter a password as
part of the two person sign-off requirement.
Second user sign-off will occur in NetTools. Refer to "Saving a batch report without a
network using USB flash drive" on page 153 and "Saving a report" on page 151.
Enter the password using the numeric keypad or keyboard.
The batch has now been approved by the current user. Time and date of approval
and user name is stored in the report for that batch.
For information on viewing reports refer to "Reports mode" on page 100.
23.4 Start Zero waste batch
i
The pump will prime and calibrate automatically and do
ongoing weight checking up to 100% so no fluid is wasted.
Disclaimer: Zero waste batch is the design and intended
purpose of the function, however, no process is
guaranteed, and loss of product could occur.
i
Option only displayed when a USB balance is detected.
Refer to "USB devices" on page 41.
Ensure balance is connected and powered on for the
duration of the batch.
i
The Flexicon Advanced Filling Kit is recommended for Zero
waste batch.